High capacity
The Rados Dosimeter Irradiator IR-2000 has been designed for calibration of the Rados TL- based dose control system. The instrument is used for sensitivity calibration of the system, for system linearity checking and for individual sensitivity calibration of the TL- material in use.
The Irradiator is microprocessor controlled and uses stepping motor for controlled irradiation of the dosimeter pellets. During the irradiation process the pellets are continuously moved past the source in order to provide precise and equal dose for each individual pellet. The drift of exposure of individual pellet in one dosimeter card is negligible.
A source of Sr90 is used to give standard doses of app. 1mSv. The irradiator may be programmed to give linearly increasing dose steps for checking the system linearity. The actual dose may be varied by repeated irradiations.
Compact and versatile
The TLD Irradiator IR-200 is a multi-purpose irradiation system with automatic drawer. The device requires only little space on the table. It is used for calibration purposes of TLD elements or TLD cards.
Different carrier discs are available for the irradiation of different TLD elements like rods, chips and discs, as well for Aluminum TLD cards and TLD Slides type RADOS. The irradiation value is defined by the number of carrier disc revolutions.
Beta Source
The IR-200 contains a 90Sr/90Y radiation source with an activity of 33 MBq.